Advances in Clinical and Experimental Medicine

2021, vol. 30, nr 11, November, p. 1205–1214

doi: 10.17219/acem/140317

Publication type: original article

Language: English

License: Creative Commons Attribution 3.0 Unported (CC BY 3.0)

Download citation:

  • BIBTEX (JabRef, Mendeley)
  • RIS (Papers, Reference Manager, RefWorks, Zotero)

Cite as:


Charong N, Kooltheat N, Plyduang T. High-sensitivity detection of clinically significant red blood cell antibodies by the column agglutination technique. Adv Clin Exp Med. 2021;30(11):1205–1214. doi:10.17219/acem/140317

High-sensitivity detection of clinically significant red blood cell antibodies by the column agglutination technique

Nurdina Charong1,A,B,C,D,E,F, Nateelak Kooltheat1,A,B,C,D,E, Thunyaluk Plyduang2,B,C,D

1 School of Allied Health Sciences, Walailak University, Nakhon Si Thammarat, Thailand

2 The Centre for Scientific and Technological Equipment, Walailak University, Nakhon Si Thammarat, Thailand

Abstract

Background. The detection of clinically significant antibodies to red blood cell antigens is important for the selection of compatible blood for patients. The conventional test tube technique (CTT) is commonly used as the gold standard test, but manual testing and visual detection of hemagglutination may produce errors. A more recently developed method, the column agglutination technique (CAT), facilitates ease of testing.

Objectives. To investigate the specificity and sensitivity of the CAT compared to the CTT for the screening of clinically significant antibodies from the Rh blood group.

Materials and methods. Standard antibodies to the Rh blood group, anti-D, -C and -E, were used as examples of clinically significant antibodies in transfusion science. The antibodies were serially diluted by two-fold, then reacted with screening cells with different antigen expression. The hemagglutination reaction was investigated using both techniques, and the grades and scores of the reactions were used to analyze the sensitivity and specificity of the CTT and CAT.

Results. The CAT had a better sensitivity than the CTT. The lowest antibody dilution of 1:8192 could be detected using CAT, while a dilution of only 1:2048 could be detected with CTT. However, the CTT and CAT were equal in specificity. The 2 techniques specifically detected all antibodies to the screening cells.

Conclusions. Both the CAT and CTT showed 100% specificity. However, the CAT exhibited more sensitivity than the CTT, and can be used in substitution of, or in parallel with, the CTT technique for red blood cell phenotyping, antibody screening, identification, and crossmatching.

Key words: column agglutination technique, conventional test tube, antibody screening, Rh antibody, anti-D

Background

Antibody screening in blood banking is used to detect unexpected antibodies to red blood cell antigens other than ABO antibodies, known as clinically significant antibodies.1 These unexpected antibodies include minor blood groups antibodies, such as anti-Rh, -MNS, -P, -Lewis, -Kell, -Kidd, -Duffy, and -Diego.2 According to the AABB standard for Blood Bank and Transfusion Service, antibody screening is performed in 2 situations. First, it is used to screen donor blood in order to prevent the passing of unexpected antibodies to patients and subsequent hemolytic transfusion reactions (HTRs). Second, antibody screening is used pre-transfusion to test a transfusion recipients sample.3 Positive antibody screening results in a donors plasma lead to the rejection of the plasma unit and, if the transfusion recipient has unexpected antibodies, antibody identification and confirmation tests are necessary to select the antigen negative blood donor for compatibility testing.4

Unexpected antibodies can be divided into 2 types according to the source of immunization. Alloantibodies are produced by patient’s immunization by foreign red blood cell antigens encountered through blood transfusion or pregnancy.5 Autoantibodies are produced after a patients immune system loses tolerance for self-antigens, which can occur with age and is associated with several diseases.6 Depending upon how immunization occurs, unexpected antibodies can also be divided into immune type and naturally occurring antibodies. Immune antibodies, such as anti-Rh, -Kidd or -Duffy, are produced after immunization through blood transfusion or pregnancy.7 These antibodies are mostly of the immunoglobulin G (IgG) type and react best at body temperature (37°C). Naturally occurring antibodies (e.g., anti-Lewis, -P) are not produced by immunization by red blood cell antigens, but by antigens that are structurally similar to those on red blood cells.8 These antibodies are mostly of the IgM type and react best at temperatures below body temperature (422°C). Clinically significant unexpected antibodies are also characterized according to their impact on donor red blood cells after blood transfusion. Highly significant IgG antibodies, such as anti-Rh, react at body temperature, and can strongly induce HTRs and produce hemolytic disease of the fetus and newborn (HDFN).9 The other type of unexpected antibodies, IgMs such as anti-I, -M and -N, react at low temperatures and normally cause a positive result in the direct antiglobulin test without hemolysis.10

Antibody screening is performed by reacting patient’s or donor’s serum/plasma with screening cells made from pooled Type O cells that have different minor red blood cell antigens.11 Unexpected antibodies in the sample will bind to the corresponding antigens on the red blood cells and directly cause hemagglutination. For the IgG class of unexpected antibodies, the addition of secondary anti-human globulins (AHG) is needed to crosslink the nearby sensitized red blood cells.

Different techniques can be used for antibody identification and an ideal technique must be able to detect low levels of clinically significant antibodies, be easy to perform, and have a short turnaround time. The conventional test tube technique (CTT) uses a glass test tube as a container for the reaction. For this test, the serum sample is mixed with screening cells, incubated at room temperature (2022°C) for the detection of IgM antibodies, and then incubated at body temperature (37°C) with AHG for the detection of IgG antibodies. The CTT is considered the gold standard for antibody screening, as it is easy to perform and it is cost effective.12 However, the grading of hemagglutination by eye and shaking the test tube by hand requires expertise, and false positive and false negatives can occur easily depending on individual techniques. The column agglutination technique (CAT) is an alternative method for antibody screening that uses a microcolumn gel card as a container for the reaction. For this technique, the serum sample and cells are transferred and mixed in a column gel card, incubated at 37°C, and spun to read the hemagglutination reaction by visual inspection or use of an automated analyzer. The CAT is easier to carry out, more accurate and more precise than the CTT technique.13 However, due to increased costs and the need for specific equipment, the CAT method is typically not used by small community hospitals.

Objectives

The aim of this study is to compare the CTT and CAT in terms of specificity, accuracy and precision for antibody screening. In order to find a suitable technique for the detection of low antibody titers, clinically significant antibodies of the Rh system (anti-D, -C, -E) were investigated.

Materials and methods

Reagents and equipment

Standard antibodies for Rh blood groups, anti-D, anti-C and anti-E, and antibody screening cells (lot: 62061) were purchased from the National Blood Centre (Thai Red Cross Society, Bangkok, Thailand). The expression of Rh antigens on the antibody screening red blood cells used in this study is shown in Table 1. The column agglutination platform and the AHG/LISS gel cards were commercially purchased (Diagnostic Grifols, Barcelona, Spain).

Antibody dilution
and antibody screening cells

Standard antibodies for Rh blood groups (anti-D, anti-C and anti-E) were two-fold diluted up to 1:16384. A total of 500 µL of the antibodies were serially diluted with an equal volume of 0.9% NaCl (normal saline solution (NSS)) using glass test tubes and a mechanical automatic pipette. The types of screening cells used were based upon expression of the Rh blood group antigens on the cell surface (Table 1). For the inspection of anti-D and anti-C, O1 or O2 screening cells were used. For the inspection of anti-E, only the O2 screening cells were used.

Antibody screening by CTT

The CTT for antibody screening was performed using the indirect antiglobulin test. First, 2 drops of diluted antibody were added to a 10 × 75 mm glass test tube. Next, 1 drop of a 3% screening cells suspension was added, and the solution was mixed and incubated at room temperature for 5 min. The tube was then centrifuged at 3000 rpm for 15 s and gently shaken for the visual inspection of hemagglutination. The solution was then further incubated at 37°C for 30 min, centrifuged for 15 s, and inspected again for hemagglutination. Incubated screening cells were then washed 3 times with NSS and the supernatant was completely discarded. Two drops of AHG (Thai Red Cross Society) were added to the tube to crosslink antibody-sensitized screening cells. The solution was then centrifuged for 15 s for the inspection of hemagglutination in the AHG phase. The reactions were graded and scored according to the standard AABB guidelines.14

Antibody screening by CAT

The antibody screening using CAT was performed according to the manufacturer’s instructions. Using an AHG/LISS microcolumn gel card, 50 µL of 1% screening cells in NSS were added, followed by 25 µL of diluted antibody. The solution was then incubated at 37°C for 15 min in a gel card incubator. The gel card was then centrifuged at 900 × g for 9 min using a specific gel card centrifuge. Hemagglutination was visually graded, scored and compared with control.

Statistical analyses

This study compared the titer of antibodies that can be detected by antibody screening. The antibody titer range was the range of the two-fold diluted antibody. The mean titer was calculated by averaging the reacted antibody titers. The mean reaction score was calculated by averaging the reaction scores of the antibody tests for the screening of D, C or E antigens on screening red blood cells.

The Gaussian distribution of all data sets from the antibody screenings were analyzed with column statistics using the D’Agostino–Pearson omnibus normality test. The antibody titers detected with the CTT and CAT were compared using paired t-tests (parametric, Gaussian distribution) or Wilcoxon matched-pairs signed-rank tests (non-parametric, non-Gaussian distribution). Correlations between the CTT and CAT were analyzed using Pearson’s correlations (parametric) or Spearmans correlations (non-parametric). The correlations between the CTT and CAT in the screening of unexpected antibodies were further analyzed with 2nd order polynomial (quadratic) to maximize the goodness of fit of the analyzed data. All statistical analyses were performed with a 95% confidence interval, p = 0.05, using GraphPad Prism v. 6.01 (GraphPad Software, San Diego, USA).

The sensitivity, specificity, accuracy, and precision of the CTT and CAT in the screening of anti-D, anti-C and anti-E were calculated using a confusion matrix.15

All methods and results are reported in agreement with the Standards for Quality Improvement Reporting Excellence (SQUIRE) guidelines.16

Results

Specificity, accuracy and precision
of CTT and CAT for antibody screening

The testing of the CTT and CAT for antibody screening was done by reacting diluted antibodies with screening cells. The screening cells expressed different antigens of the Rh blood group, with O1 cells expressing only D and C antigens and O2 cells expressing D, C and E antigens (Table 1). The specificity of the antibody screening techniques was analyzed using the reaction between the screening cells and specific antibodies. The CTT and CAT both gave positive results for reactions between the O1 or O2 cells and anti-D and anti-C (Table 2, Table 3). The 2 techniques also gave negative results for the reaction of O1 and positive results for the reaction of O2, when testing with anti-E (Table 4). These results show that both the CTT and CAT have equal specificity in the screening of clinically significant antibodies to D, C and E antigens. Due to the use of E antigen-negative cells in the screening test, the specificity of anti-E was also calculated (Table 5). The precision of the antibody screening results performed by the CAT was consistent when compared to the CTT and showed a high accuracy for all tested antibodies (Table 5).

Sensitivity of CTT and CAT
for antibody screening

The screening of antibodies using CTT and CAT was performed according to the AABB standard protocols.4 The reaction of hemagglutination was visually graded as negative (0), weakly positive (w+) or positive (1+, 2+, 3+, 4+). The same reaction was also scored as negative (2), weakly positive (4) or positive (5, 8, 10, 12).

The CTT detected anti-D, anti-C and anti-E at the lowest dilutions of 1:2048, 1:1024 and 1:2048, respectively. The reaction scores for the lowest dilutions of anti-D, anti-C and anti-E were 5, 4 and 4, respectively. The total scores for hemagglutination for anti-D, anti-C and anti-E were 227.0 ±2.0, 190.70 ±5.70 and 138.30 ±2.50, respectively (Table 2, Table 3, Table 4).

The CAT detected anti-D, anti-C and anti-E at the lowest dilutions of 1:8192, 1:1024 and 1:4096, respectively. Reaction scores for the lowest dilutions of anti-D, anti-C and anti-E were all at 5. The total scores for hemagglutination for anti-D, anti-C and anti-E were 288.0 ±2.0, 220.3 ±7.4 and 152.±4.9, respectively (Table 2, Table 3, Table 4).

Column statistics were calculated and the Gaussian distribution of all data sets was tested. Data from the anti-D screening with CTT, and the anti-E screening with CTT and CAT, were normally distributed (Table 1). Data from the anti-D screening with CAT, and the anti-C screening with CTT and CAT, were not normally distributed (Table 1). Thus, parametric tests were used for data sets from the anti-E screenings, while non-parametric tests were applied to data sets from anti-D and anti-C screenings.

The results indicated that CAT had a significantly higher sensitivity than CTT (0.0313 compared to 0.0285, respectively, p = 0.0078; Figure 1A–Figure 3A).

Titer range, mean titer and mean
reaction scores of antibodies detected using CTT and CAT

The sensitivity of the CTT and CAT was compared by examining the mean reaction scores, titer ranges and the mean titers of detectable antibodies. The CTT had mean reaction scores for the screening of anti-D, anti-C and anti-E of 227.0 ±2.0, 190.70 ±5.70 and 138.30 ±2.50, respectively. The antibody titer ranges detected using CTT for anti-D, anti-C and anti-E were all at 016387 (Table 2, Table 3, Table 4). The CAT had mean reaction scores in the screening of anti-D, anti-C and anti-E of 288.0 ±2.0, 220.3 ±7.4 and 152.7 ±4.9, respectively. The antibody titer ranges detected with CTT for anti-D, anti-C and anti-E were all at 016387 (Table 2, Table 3, Table 4). The column agglutination technique had a better sensitivity in antibody screening than CAT (Figure 1B–Figure 3B).

Differences between CTT and CAT in screening for unexpected clinically significant Rh antibodies

Differences in the screening of anti-D, anti-C and anti-E were observed, as CAT has a higher sensitivity than CTT. The screening of anti-D using CAT was significantly different from screening with CTT, with a p-value of 0.0078, a Spearman’s rank correlation coefficient of 0.8711 and a R-square of 0.8541 (Table 6, Figure 1C). The screening of anti-C using CAT was significantly different from screening with CTT, with a p-value of 0.0313, a Spearman’s rank correlation coefficient of 0.9988 and a R-square of 0.9913 (Table 6, Figure 2C). The screening of anti-E using CAT was significantly different from screening with CTT with a p-value of 0.0285, a Pearson’s correlation coefficient of 0.9234 and a R-square of 0.8919 (Table 6, Figure 3C).

Discussion

The CTT is considered the gold standard for tests in the blood bank laboratory, including for ABO, Rh typing, antibody screening, antibody identification, and crossmatching.5 This technique uses only basic equipment, is cost-effective and gives significant results in the reading of the reaction. In addition, antibodies of both the IgM and IgG classes can be detected in a single tube at different temperature phases. However, the CTT technique depends on manual testing and reading by the eyes, which may be subject to human error, especially under a high workload. Moreover, the techniques used for tube shaking to resuspend the red cell pellet for observing agglutination may vary among laboratory staff.

In 1985, Lapierre et al. developed a new technique, CAT, for detection of the reaction between red blood cell antigens and antibodies for red blood cell phenotyping.17 This method uses a microcolumn packed with gel particles, specific antibodies, LISS buffer, and Coombs/AHG reagent. The Food and Drug Administration (FDA) issued the first gel technology license in 1994 to Micro Typing System under the product name ID-MTS Card. The implementation of this technique in hospital laboratories and transfusion services within 5 years after the release of the first report indicated widespread satisfaction with the test.18, 19 The CAT is particularly useful as it can eliminate many of the variables associated with manual testing and provides stable endpoint agglutination. The CAT also minimizes sample preparation as can be performed using an automated analyzer, and the results can be repeatedly verified.5

The implementation of the CAT for pretransfusion testing was also reported in a community hospital, and indicated an increase in the overall number of antibodies and clinically significant antibodies detected with CAT, as compared to CTT, without significant differences in non-specific reactions.20 Earlier studies also compared the CAT with the CTT in terms of ABO, Rh phenotyping, antibody screening, and crossmatching. A strong correlation between the CAT and CTT results for both anti-A and anti-B IgG titers was reported, but these authors suggested that the relationship between CTT and CAT for the titration of other kinds of antibodies should be further analyzed.21 Another study also indicated that CAT results showed a high degree of concordance with the CTT, suggesting that the CAT could be brought into routine use. However, proper training and standardization are needed prior to the use of the CAT.22 In addition, the costs associated with the CAT are increased compared to the CTT. In the current study, the cost per test for these 2 methods was calculated for a comparison, based on the estimated number of tests per day (10 tests/day) performed by a blood bank in a general or community hospital (Table 7). Interestingly, the estimated cost for CAT is 100% higher than that for CTT. Thus, CAT should not be recommended for routine use in the laboratory. However, using the CAT in parallel with the CTT for specific cases, as indicated in Figure 4, is suggested to reduce the adverse effects of transfusion due to the secondary immunization of alloantibody (e.g., delayed hemolytic transfusion reaction (DHTR)).

Although many previous studies have reported on the advantages of using the CAT for detecting unexpected antibodies in terms of sensitivity and specificity, our study showed gaps in the antibody titers that can be detected with these methods. Frequently detectable antibodies to the Rh blood group antigens, anti-D, anti-C and anti-E, were used as model antibodies, which can be applied in the investigation of DHTR and HDFN in real-life situations. While the CAT was significantly comparable to the gold standard technique CTT (p < 0.01), the CAT had a higher sensitivity than the standard CTT, and the mean titer of CAT in the detection of anti-D, anti-C and anti-E was significantly higher than CTT (p < 0.01).

Thus, the use of CAT instead of CTT may be better in detecting lower-titer antibodies, especially for the clinically significant antibodies. In addition, to prevent DHTRs, CAT should be carried out for the detection of low-titer antibodies that have been previously detected in a patient, but where no agglutination appeared with the CTT. Nevertheless, IgM class unexpected antibodies cannot be detected using CAT, as this technique does not have the ability to detect antibodies at room temperature. Therefore, in the case of cold alloantibody detection, the use of CAT in pre-transfusion compatibility testing or crossmatching should be performed in parallel with CTT or using the neutral Gel card. Column agglutination technique has been reported to estimate anti A/B titer in an ABO incompatible kidney transplant, which indicated a sensitivity approx. 2.5-fold higher than CTT.23 However, CAT was not suggested to determine a critical titer for anti-D titration, since it showed significantly higher titers than the CTT in all samples studied.24

Limitations

The current study indicates the superior sensitivity of the CAT technique with a wide-gap titer in the detection of commercially available antibodies to the Rh system, compared with the CTT. However, applying the CAT to detect the presence of significant unexpected antibodies in clinical samples is needed to confirm the advantages of the CAT over the CTT. The simpler CAT technique minimizes turnaround time and workload while maintaining standard operation and results, and can easily be performed by blood bank laboratory technologists.

Conclusions

The CAT can be used in substitution of, or in parallel with, the CTT for the screening of clinically significant antibodies (especially those with a low titer) to avoid secondary immunization of antibodies to minor red blood cell antigens after transfusion.

Tables


Table 1. Expression of Rh antigens on antibody screening red blood cells

Screening cells

Rh

Wiener

D

C

E

c

e

O1

R1R1

+

+

0

0

+

O2

R1R2

+

+

+

+

+

Table 2. Reaction gradings and scores of anti-D screening using CTT and CAT

Anti-D

Dilution (titer range)

conventional test tube technique

(CTT)

column agglutination technique

(CAT)

reaction grading

score

reaction grading

score

O1

O2

O1

O2

O1

O2

O1

O2

1st

2nd

3rd

1st

2nd

3rd

1st

2nd

3rd

1st

2nd

3rd

1st

2nd

3rd

1st

2nd

3rd

1st

2nd

3rd

1st

2nd

3rd

1:2

4+

4+

4+

4+

4+

4+

12

12

12

12

12

12

4+

4+

4+

4+

4+

4+

12

12

12

12

12

12

1:4

4+

4+

4+

4+

4+

4+

12

12

12

12

12

12

4+

4+

4+

4+

4+

4+

12

12

12

12

12

12

1:8

4+

4+

4+

4+

4+

4+

12

12

12

12

12

12

4+

4+

4+

4+

4+

4+

12

12

12

12

12

12

1:16

4+

4+

4+

4+

4+

4+

12

12

12

12

12

12

4+

4+

4+

4+

4+

4+

12

12

12

12

12

12

1:32

4+

4+

4+

4+

4+

4+

12

12

12

12

12

12

4+

4+

4+

4+

4+

4+

12

12

12

12

12

12

1:64

3+

4+

3+

3+

4+

3+

10

12

10

10

12

10

4+

4+

4+

4+

4+

4+

12

12

12

12

12

12

1:128

2+

2+

2+

3+

4+

3+

8

8

8

10

12

10

4+

4+

4+

4+

4+

4+

12

12

12

12

12

12

1:256

2+

2+

2+

2+

4+

3+

8

8

8

8

12

10

4+

4+

4+

4+

4+

4+

12

12

12

12

12

12

1:512

2+

1+

2+

2+

2+

2+

8

5

8

8

8

8

4+

4+

4+

4+

4+

4+

12

12

12

12

12

12

1:1024

2+

1+

2+

1+

1+

1+

8

5

8

5

5

5

3+

4+

3+

3+

4+

4+

10

12

10

10

12

12

1:2048

1+

1+

1+

1+

1+

1+

5

5

5

5

5

5

3+

3+

3+

3+

3+

3+

10

10

10

10

10

10

1:4096

0

0

0

0

0

0

2

2

2

2

2

2

2+

2+

2+

2+

2+

2+

8

8

8

8

8

8

1:8192

0

0

0

0

0

0

2

2

2

2

2

2

1+

1+

1+

1+

1+

1+

5

5

5

5

5

5

1:16384

0

0

0

0

0

0

2

2

2

2

2

2

0

0

0

0

0

0

2

2

2

2

2

2

Sum

113

109

113

112

120

114

Sum

143

145

143

143

145

145

Average

111.7 ±2.3

115.4 ±4.2

Average

143.7 ±1.2

144.3 ±1.2

Total (mean reaction score)

227.0 ±2.0

Total (mean reaction score)

288.0 ±2.0

Table 3. Reaction gradings and scores of anti-C screening using CTT and CAT

Anti-C

Dilution (titer range)

conventional test tube technique

(CTT)

column agglutination technique

(CAT)

reaction grading

score

reaction grading

score

O1

O2

O1

O2

O1

O2

O1

O2

1st

2nd

3rd

1st

2nd

3rd

1st

2nd

3rd

1st

2nd

3rd

1st

2nd

3rd

1st

2nd

3rd

1st

2nd

3rd

1st

2nd

3rd

1:2

4+

4+

4+

4+

4+

4+

12

12

12

12

12

12

4+

4+

4+

4+

4+

4+

12

12

12

12

12

12

1:4

4+

4+

4+

4+

4+

4+

12

12

12

12

12

12

4+

4+

4+

4+

4+

4+

12

12

12

12

12

12

1:8

4+

4+

4+

4+

4+

4+

12

12

12

12

12

12

4+

4+

4+

4+

4+

4+

12

12

12

12

12

12

1:16

4+

4+

4+

4+

4+

4+

12

12

12

12

12

12

4+

4+

4+

4+

4+

4+

12

12

12

12

12

12

1:32

4+

3+

3+

2+

2+

2+

12

10

10

8

8

8

4+

4+

4+

4+

3+

4+

12

12

12

12

10

12

1:64

3+

4+

3+

2+

1+

2+

10

12

10

10

5

8

4+

4+

4+

4+

3+

4+

12

12

12

12

10

12

1:128

2+

2+

2+

1+

1+

1+

8

8

8

5

5

5

4+

4+

4+

2+

3+

2+

12

12

12

8

10

8

1:256

2+

2+

1+

1+

1+

1+

8

8

5

5

5

5

3+

4+

3+

1+

2+

2+

10

12

10

5

8

8

1:512

1+

2+

1+

w+

1+

0

5

8

5

4

5

2

2+

3+

3+

0

1+

0

8

10

10

2

5

2

1:1024

w+

1+

w+

w+

0

0

4

5

4

4

2

2

1+

2+

2+

0

1+

0

5

8

8

2

5

5

1:2048

0

0

0

0

0

0

2

2

2

2

2

2

0

0

0

0

0

0

2

2

2

2

2

2

1:4096

0

0

0

0

0

0

2

2

2

2

2

2

0

0

0

0

0

0

2

2

2

2

2

2

1:8192

0

0

0

0

0

0

2

2

2

2

2

2

0

0

0

0

0

0

2

2

2

2

2

2

1:16384

0

0

0

0

0

0

2

2

2

2

2

2

0

0

0

0

0

0

2

2

2

2

2

2

Sum

103

107

98

92

86

86

Sum

115

122

120

97

104

103

Average

102.7 ±4.5

88.0 ±3.5

Average

119.0 ±3.6

101.3 ±3.8

Total (mean reaction score)

190.7 ±5.7

Total (mean reaction score)

220.3 ±7.4

w+ – weakly positive.
Table 4. Reaction gradings and scores of anti-E screening using CTT and CAT

Anti-E

Dilution (titer range)

conventional test tube technique

(CTT)

column agglutination technique

(CAT)

reaction grading

score

reaction grading

score

O1

O2

O1

O2

O1

O2

O1

O2

1st

2nd

3rd

1st

2nd

3rd

1st

2nd

3rd

1st

2nd

3rd

1st

2nd

3rd

1st

2nd

3rd

1st

2nd

3rd

1st

2nd

3rd

1:2

0

0

0

4+

4+

4+

2

2

2

12

12

12

0

0

0

4+

4+

4+

2

2

2

12

12

12

1:4

0

0

0

4+

4+

4+

2

2

2

12

12

12

0

0

0

4+

4+

4+

2

2

2

12

12

12

1:8

0

0

0

4+

4+

4+

2

2

2

12

12

12

0

0

0

4+

4+

4+

2

2

2

12

12

12

1:16

0

0

0

4+

4+

4+

2

2

2

12

12

12

0

0

0

4+

4+

4+

2

2

2

12

12

12

1:32

0

0

0

4+

4+

3+

2

2

2

12

12

10

0

0

0

4+

4+

4+

2

2

2

12

12

12

1:64

0

0

0

4+

3+

3+

2

2

2

12

10

10

0

0

0

4+

4+

3+

2

2

2

12

12

10

1:128

0

0

0

3+

2+

2+

2

2

2

10

8

8

0

0

0

3+

4+

3+

2

2

2

10

12

10

1:256

0

0

0

2+

2+

2+

2

2

2

8

8

8

0

0

0

3+

4+

3+

2

2

2

10

12

10

1:512

0

0

0

1+

2+

2+

2

2

2

5

8

8

0

0

0

2+

3+

2+

2

2

2

8

10

18

1:1024

0

0

0

1+

2+

1+

2

2

2

5

8

5

0

0

0

1+

2+

1+

2

2

2

5

8

5

1:2048

0

0

0

w+

1+

1+

2

2

2

4

5

5

0

0

0

1+

1+

1+

2

2

2

5

5

5

1:4096

0

0

0

0

0

0

2

2

2

2

2

2

0

0

0

1+

1+

1+

2

2

2

5

5

5

1:8192

0

0

0

0

0

0

2

2

2

2

2

2

0

0

0

0

0

0

2

2

2

2

2

2

1:16384

0

0

0

0

0

0

2

2

2

2

2

2

0

0

0

0

0

0

2

2

2

2

2

2

Sum

28

28

28

110

113

108

Sum

28

28

28

119

128

127

Average

28.0 ±0.0

115.3 ±2.5

Average

28.0 ±0.0

124.7 ±4.9

Total (mean reaction score)

138.3 ±2.5

Total (mean reaction score)

152.±4.9

w+ – weakly positive.
Table 5. Sensitivity, specificity, accuracy, and precision of CTT and CAT in the screening of anti-D, anti-C and anti-E

Measure

Conventional test tube technique (CTT)

Column agglutination technique

(CAT)

anti-D

anti-C

anti-E

anti-D

anti-C

anti-E

Sensitivity

0.7857

0.6786

0.7857

0.9286

0.6667

0.8571

Specificity

1

1

Accuracy

0.8757

0.6786

0.8929

0.9286

0.6667

0.9286

Precision

1

1

1

1

1

1

Matthews correlation coefficient

0.8044

0.866

Table S1. Statistical analyses of reaction score in screening of anti-D, anti-C and anti-E using conventional tube test technique (CTT) and column agglutination technique (CAT)

Parameter

Scrn. of anti-D

Scrn. of anti-C

Scrn. of anti-E

CTT

CAT

CTT

CAT

CTT

CAT

Mean

8.107

10.29

6.810

7.869

7.881

8.905

Standard deviation

4.029

3.148

4.186

4.389

3.986

3.960

Standard error of mean

1.077

0.8414

1.119

1.173

1.065

1.058

Lower 95% CI of mean

5.781

8.468

4.392

5.335

5.579

6.618

Upper 95% CI of mean

10.43

12.10

9.227

10.40

10.18

11.19

D’Agostino–Pearson omnibus normality test

K2

3.087

13.12

6.095

6.398

3.269

3.136

p-value

0.2136

0.0014

0.0475

0.0408

0.1950

0.2085

Passed normality test? (alpha = 0.05)

yes

no

no

no

yes

yes

p-value summary

ns

**

*

*

ns

ns

Coefficient of variation

49.69%

30.61%

61.48%

55.77%

50.58%

44.47%

* p < 0.05; ** p < 0.01; ns – not significant; Scrn. – screening.
Table 6. Differences between conventional test tube technique (CTT) and column agglutination technique (CAT) in screening of unexpected clinically significant Rh antibodies

Parameter

Screening of unexpected antibody

anti-D

anti-C

anti-E

Pared t-test

p-value

0.0285

t-statistic

2.463

Degrees of freedom (df)

13

Mean of differences

–0.5119

95% confidence interval

–0.9608 to –0.06297

R-square

0.3182

Pearson correlation coefficient (r)

0.9234

p-value

<0.0001

Wilcoxon matched-pairs signed-rank test

p-value

0.0078

0.0313

Sum of positive

0.0

0.0

Negative ranks

–84.0

–69.0

Sum of signed ranks

–84.0

–69.0

Median of differences

–2.0

0.0

98.71% confidence interval

–5.0 to 0.0

–2.5 to 0.0

Spearman’s rank correlation coefficient (Rs)

0.8711

0.9988

p-value

<0.0001

<0.0001

Table 7. Cost comparison analysis of antibody screening by conventional tube test technique (CTT) and column agglutination technique (CAT)

Equipment/reagent

CTT

CAT

system

cost*

[USD/year]

system

cost*

[USD/year]

Reaction container

10 × 75 Glass Test Tube (reusable)

(PYREX VISTA, Corning Inc., Glendale, USA)

(1000 tubes)

380

(0.38/Tube)

AHG/LISS Microcolumn (Disposable)
(8 reactions/card, 50 cards/box)

(DG Gel Coombs, Grifols, S.A., Barcelona, Spain)

(19 boxes)

5947

(313/Box)

Screening cell

O1, O2 (25 µL/test, 10 mL/bottle/month)

(National Blood Centre, Bangkok, Thailand)

(12 bottles O1, 12 bottle O2)

28.8

(1.2/Bottle)

O1, O2 (25 µL/test, 10 mL/bottle/month)

(National Blood Centre, Bangkok, Thailand)

(12 bottles O1, 12 bottle O2)

28.8

(1.2/bottle)

Anti-human globulin reagent

AHG/Coomb’s (50 µL/test, 10 mL/bottle)

(National Blood Centre, Bangkok, Thailand)

(37 bottles)

118.4

(3.2/Bottle)

Low ionic strength solution

LISS (100 µL/test, 10 mL/bottle)

(National Blood Centre, Bangkok, Thailand)

(74 bottles)

236.8

(3.2/Bottle)

Incubator

Heat Block

(DVT-2B, BioVision Inc., Milpitas, USA)

360

Microcolumn Card Incubator

(DG Therm, Grifols, S.A., Barcelona, Spain)

1625

Centrifuge

Serofuge

(EBA 280 Serology Decanting Package 5, Andreas Hettich GmbH & Co. KG, Tuttlingen, Germany)

4481

Microcolumn Centrifuge

(DG Spin, Grifols, S.A., Barcelona, Spain)

5063

Tube cleaning cost

every week (48 ×10 USD)

480

Total

7300 tests/year

6085

7300 tests/year

12663.8

Cost

per 1 test

0.83

per 1 test

1.73

Difference

CTT compared to CAT

208.12%

CAT compared to CTT

48.05%

Cost analyzed from product supplied in Thailand. * 10 tests of antibody screening (O1, O2) per day, 7300 tests for 1 year.

Figures


Fig. 1. Sensitivity comparison of the conventional test tube (CTT) and column agglutination techniques (CAT) in the screening of anti-D. Anti-D was two-fold diluted from 1:2 to 1:16348 and then screened with O1 and O2 cells using CTT and CAT. The hemagglutination of each reaction was scored. A. Sensitivity of CTT and CAT was compared with Wilcoxon matched-pairs signed-rank test; B. Antibody titers ranging from 1-16384 for anti-D detection using CTT and CAT were used; C. Correlation between CTT and CAT was analyzed with Spearman’s correlation and non-linear curve fitting
Fig. 2. Sensitivity comparison of the conventional test tube (CTT) and column agglutination techniques (CAT) in the screening of anti-C. Anti-C was two-fold diluted from 1:2 to 1:16348 then screened with O1 and O2 cells using CTT and CAT. The hemagglutination of each reaction was scored. A. Sensitivity of CTT and CAT was compared with Wilcoxon matched-pairs signed-rank test; B. Antibody titers ranging from 1 to 16384 for anti-C detection using CTT and CAT were used; C. Correlation between CTT and CAT was analyzed with Spearman’s correlation and non-linear curve fitting
Fig. 3. Sensitivity comparison of conventional test tube (CTT) and column agglutination techniques (CAT) in the screening of anti-E. Anti-E was two-fold diluted from 1:2 to 1:16348 then screened with O1 and O2 cells using CTT and CAT. The hemagglutination of each reaction was scored. A. Sensitivity of CTT and CAT was compared by matched-pairs t-test; B. Antibody titers ranging from 1 to 16384 for anti-E detection using CTT and CAT were used; C. Correlation between CTT and CAT was analyzed with Pearson’s correlation and non-linear curve fitting
Fig. 4. Suggested guidelines for using column agglutination technique (CAT) in parallel with conventional tube test technique (CTT) in specific case of blood requesting

References (24)

  1. Fung MK, Grossman BJ, Hillyer CD, Westhoff CM, eds. Technical Manual. 18th ed. Bethesda, USA: AABB; 2014:391-396.
  2. Giblett ER. Blood group alloantibodies: An assessment of some laboratory practices. Transfusion. 1997;17(4):299–308. doi:10.1046/j.1537-2995.1977.17477216857.x
  3. Nance ST. Red cell antibody detection by serology. ISBT Sci Ser. 2015;10(Suppl 1):1–4:225–231. doi:10.1111/voxs.12128
  4. Walker PS, Hamilton JR. Identification of antibodies to red cell antigens. In: Fung MK, Grossman BJ, Hillyer CD, Westhoff CM, eds. Technical Manual. 18th ed. Bethesda, USA: AABB; 2014:391–421.
  5. Avent ND. Transfusion and Transplantation Science. Oxford, UK: Oxford University Press; 2018.
  6. Tomer Y, Shoenfeld Y. Ageing and autoantibodies. Autoimmunity. 1988;1(2):141–149. doi:10.3109/08916938809001927
  7. Daniels G, Poole J, de Silva M, et al. The clinical significance of blood group antibodies. Transfus Med. 2002;12(5):287–295. doi:10.1046/j.1365-3148.2002.00399.x
  8. Quinley ED. Immunohematology Principles and Practice. Philadelphia, USA: Lippincott Williams and Wilkins; 2011.
  9. Urbaniak SJ, Greiss MA. RhD haemolytic disease of the fetus and the newborn. Blood Rev. 2000;14(1):44–61. doi:10.1054/blre.1999.0123
  10. Poole J, Daniels G. Blood group antibodies and their significance in transfusion medicine. Transfus Med Rev. 2007;21(1):58–71. doi:10.1016/j.tmrv.2006.08.003
  11. Downes KA, Shulman IA. Antigen and antibody testing. In: Fung MK, Grossman BJ, Hillyer CD, Westhoff CM, eds. Technical Manual. 18th ed. Bethesda, USA: AABB; 2014:367–384.
  12. Rumsey DH, Ciesielski DJ. New protocols in serologic testing: A review of techniques to meet today’s challenges. Immunohematology. 2000;16(4):131–137. PMID:15373603
  13. Langston MM, Procter JL, Cipolone KM, Stroncek DF. Evaluation of the gel system for ABO grouping and D typing. Transfusion. 1999;39(3):300–305. doi:10.1046/j.1537-2995.1999.39399219288.x
  14. Roback JD, Grossman BJ, Harris T, Hillyer CD, eds. Technical Manual. 17th ed. Bethesda,USA: AABB; 2011:873-874.
  15. Chicco D, Tötsch N, Jurman G. The Matthews correlation coefficient (MCC) is more reliable than balanced accuracy, bookmaker informedness, and markedness in two-class confusion matrix evaluation. BioData Min. 2021;14(1):13. doi:10.1186/s13040-021-00244-z
  16. Ogrinc G, Davies L, Goodman D, et al. SQUIRE 2.0 (Standards for Quality Improvement Reporting Excellence): Revised publication guidelines from a detailed consensus process. BMJ Qual Saf. 2016;25:986–992. doi:10.1016/j.jamda.2021.06.006
  17. Lapierre Y, Rigal D, Adam J, et al. The gel test: A new way to detect red cell antigen-antibody reactions. Transfusion. 1990;30(2):109–113. doi:10.1046/j.1537-2995.1990.30290162894.x
  18. Cheng G, Chan A, Chun M, et al. Evaluation of the gel test for antibody screening in a tertiary hospital in Hong Kong: Insensitivity for some cold antibodies that are reactive at 37°C by conventional indirect antiglobulin tests. Clin Lab Haematol. 1995;17(1):81–84. doi:10.1111/j.1365-2257.1995.tb00323.x
  19. Chan HF, Wong CH, Chui L, et al. The impact of a gel system on routine work in a general hospital blood bank. Immunohematology. 1996;12(1):30–32. PMID:15387759
  20. Delaflor-Weiss E, Chizhevsky V. Implementation of gel testing for antibody screening and identification in a community hospital: A 3-year experience. Lab Med. 2005;36(8):489–492. doi:10.1309/JAP6EC69BAAUG9B3
  21. Cheng D, Hao Y. Comparative evaluation of the microcolumn gel card test and the conventional tube test for measurement of titres of immunoglobulin G antibodies to blood group A and blood group B. J Int Med Res. 2011;39(3):934–943. doi:10.1177/147323001103900328
  22. Bhagwat SN, Sharma JH, Jose J, et al. Comparison between conventional and automated techniques for blood grouping and crossmatching: Experience from a tertiary care centre. J Lab Physicians. 2015;7(2):96–102. doi:10.4103/0974-2727.163130
  23. Bhangale A, Pathak A, Pawar S, et al. Comparison of antibody titers using conventional tube technique versus column agglutination technique in ABO blood group incompatible renal transplant. Asian J Transfus Sci. 2017;11(2):131–134. doi:10.4103/0973-6247.214343
  24. Novaretti MCZ, Jens E, Pagliarini T, et al. Comparison of conventional tube test with diamed gel microcolumn assay for anti-D titration. Clin Lab Haematol. 2003;25(5):311–315. doi:10.1046/j.1365-2257.2003.00540.x