Advances in Clinical and Experimental Medicine

2018, vol. 27, nr 4, April, p. 481–485

doi: 10.17219/acem/68559

Publication type: original article

Language: English

Download citation:

  • BIBTEX (JabRef, Mendeley)
  • RIS (Papers, Reference Manager, RefWorks, Zotero)

A comparison of intraocular pressure values obtained using a Goldmann applanation tonometer and a handheld version of applanation resonance tonometer: A preliminary report

Małgorzata Mulak1,A,B,C,D,E,F, Wojciech A. Czak1,A,B,C,D, Małgorzata Mimier1,D,E, Radosław Kaczmarek1,C,E

1 Department and Clinic of Ophthalmology, Wroclaw Medical University, Poland

Abstract

Background. Despite the development of various methods of intraocular pressure (IOP) measurement, Goldmann applanation tonometry (GAT) is still the most popular. The measurement using GAT depends on the biomechanical properties of the cornea, such as the thickness, the radius of curvature, as well as the amount of the fluorescein used.
Objectives. The aim of the study was to compare IOP values measured by GAT with those measured by applanation resonance tonometry (ART).
Material and Methods. A total of 47 patients (94 eyes), including 28 patients with primary open-angle glaucoma (POAG) and 19 subjects from the control group, were examined at the Glaucoma Outpatient Clinic of the Department and Clinic of Ophthalmology at Wroclaw Medical University (Poland). The measurements of IOP were performed using GAT and a handheld version of ART. Also, the central corneal thickness (CCT) of all patients was measured.
Results. The study showed that the IOP values measured by both tonometers were comparable, but ARTacquired values were higher than GAT-obtained values both in the glaucomatous group and in the control group. CCT had little impact on mean IOP difference between GATand ART-obtained values.
Conclusion. Applanation resonance tonometry is a precise method of IOP measurement and is less affected by biomechanical properties of the cornea than GAT. Our results show that ART is a new, promising, comfortable for both patients and doctors method of IOP measurement, which, in the future, can replace GAT.

Key words

comparison, glaucoma, ART, tonometry, GAT

References (18)

  1. Terminology and Guidelines for Glaucoma. 4th ed. Savona, Italy: European Glaucoma Society; 2014.
  2. Goldmann H, Schmidt T. Applanation tonometry. (in German) Ophthalmologica. 1957;134(4):221–242.
  3. Clement CI, Parker DG, Goldberg I. Intraocular pressure measurement in a patient with a thin, thick or abnormal cornea. Open Ophthalmol J. 2016;10:35–43.
  4. Cacho I, Sanchez-Naves J, Batres L, Pintor J, Carracedo G. Comparison of intraocular pressure before and after laser in situ keratomileusis refractive surgery measured with Perkins tonometry, noncontact tonometry and transpalpebral tonometry. J Ophthalmol. 2015;2015:683895.
  5. Eklund A, Hallberg P, Lindén C, Lindahl OA. An applanation resonator sensor for measuring intraocular pressure using combined continuous force and area measurement. Invest Ophthalmol Vis Sci. 2003;44(7):3017–3024.
  6. Hallberg P, Lindén C, Lindahl OA, Bäcklund T, Eklund A. Applanation resonance tonometry for intraocular pressure in humans. Physiol Meas. 2004;25(4):1053–1065.
  7. Jóhannesson G, Hallberg P, Eklund A, Lindén C. Introduction and clinical evaluation of servo-controlled applanation resonance tonometry. Acta Ophthalmol. 2012;90(7):677–682.
  8. Ottobelli L, Fogagnolo P, Frezzotti P, et al. Repeatability and reproducibility of applanation resonance tonometry: A cross-sectional study. BMC Ophthalmol. 2015;15:36.
  9. Mulak M, Czak W, Grobarek B, Borwińska M, Misiuk-Hojło M. Intraocular pressure measurements obtained using Goldmann applanation tonometry (GAT) and applanation resonance tonometry (ART) – a comparison. Magazyn Lekarza Okulisty. 2015;9(3):140–145.
  10. Hallberg P, Santala K, Lindén C, Lindahl OA, Eklund A. Comparison of Goldmann applanation and applanation resonance tonometry in a biomicroscope-based in vitro porcine eye model. J Med Eng Technol. 2006;30(6):345–352.
  11. Hallberg P, Lindén C, Bäcklund T, Eklund A. Symmetric sensor for applanation resonance tomometry of the eye. Med Biol Eng Comput. 2006;44(1–2):54–60.
  12. Hallberg P, Eklund A, Bäcklund T, Lindén C. Clinical evaluation of applanation resonance tonometry: A comparison with Goldmann applanation tonometry. J Glaucoma. 2007;16(1):88–93.
  13. Salvetat ML, Zeppieri M, Tosoni C, Brusini P. Repeatability and accuracy of applanation resonance tonometry in healthy subjects and patients with glaucoma. Acta Ophthalmol. 2014;92(1):66–73.
  14. Jóhannesson G, Hallberg P, Eklund A, Koskela T, Lindén C. Change in intraocular pressure measurement after myopic LASEK: A study evaluating Goldmann, Pascal and applanation resonance tonometry. J Glaucoma. 2012;21(4):255–259.
  15. Jóhannesson G, Hallberg P, Eklund A, Koskela T, Lindén C. Change in intraocular pressure measurement 2 years after myopic laser-assisted subepithelial keratectomy. J Cataract Refract Surg. 2012;38(9):1637–1642.
  16. Jóhannesson G, Hallberg P, Ambarki K, Eklund A, Lindén C. Age-dependency of ocular parameters: A cross-sectional study of young and elderly healthy subjects. Graefes Arch Clin Exp Ophthalmol. 2015;253(11):1979–1983.
  17. Beckman RJ, Behndig A, Hallberg P, Linden C. Increased corneal hysteresis after corneal collage cross linking: A study based on applanation resonance technology. JAMA Ophthalmol. 2014;132(12):1426–1432.
  18. Jóhannesson G, Hallberg P, Eklund A, Behndig A, Lindén C. Effects of topical anaesthetics and repeated tonometry on intraocular pressure. Acta Ophthalmol. 2014;92(2):111–115.